Molecularly Imprinted Polypyrrole-Modified Screen-Printed Electrode for Dopamine Determination
Articolo
Data di Pubblicazione:
2024
Abstract:
: This paper introduces a quantitative method for dopamine determination. The method is based on a molecularly imprinted polypyrrole (e-MIP)-modified screen-printed electrode, with differential pulse voltammetry (DPV) as the chosen measurement technique. The dopamine molecules are efficiently entrapped in the polymeric film, creating recognition cavities. A comparison with bare and non-imprinted polypyrrole-modified electrodes clearly demonstrates the superior sensitivity, selectivity, and reproducibility of the e-MIP-based one; indeed, a sensitivity of 0.078 µA µM-1, a detection limit (LOD) of 0.8 µM, a linear range between 0.8 and 45 µM and a dynamic range of up to 350 µM are achieved. The method was successfully tested on fortified synthetic and human urine samples to underline its applicability as a screening method for biomedical tests.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
differential pulse voltammetry; dopamine; electropolymerized molecularly imprinted polypyrrole; molecularly imprinted polymers; polypyrrole-based sensors; screen-printed electrodes
Elenco autori:
Merli, Daniele; Cutaia, Alessandra; Hallulli, Ines; Bonanni, Alessandra; Alberti, Giancarla
Link alla scheda completa: