  1. Outputs

Absolute thermal neutron fluence determination by thin film of natural uranium

Academic Article
Publication Date:
1995
abstract:
An absolute monitor of thermal neutron fluence based on the 235U induced fission was developed. This monitor is constituted by a solid state nuclear track detector juxtaposed to a natural uranium film with a negligible self-absorption to fission fragments. In order to perform the calibration of the films, the alpha-activity was measured by using nuclear emulsions. The preparation, calibration and employment procedures of this monitor are presented.
Iris type:
1.1 Articolo in rivista
List of contributors:
G., Bigazzi; J. C. Hadler, N.; P. J., Iunes; Oddone, Massimo; R. S., Paulo; A. Zuñiga, G.
Handle:
https://iris.unipv.it/handle/11571/492453
Published in:
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT
Journal