Behaviour in magnetic fields of fast conventional and fine-mesh photomultipliers

Academic Article
Publication Date:
2012
abstract:
The performance of both conventional and fine-mesh Hamamatsu photomultipliers has been measured inside moderate magnetic fields. This has allowed the test of effective shielding solutions for photomultipliers, to be used in time-of-flight detectors based on scintillation counters. Both signal amplitude reduction and deterioration of the timing properties inside magnetic fields have been investigated.
Iris type:
1.1 Articolo in rivista
Keywords:
Time of flight detectors; PHOTOMULTIPLIERS
List of contributors:
M., Bonesini; R., Bertoni; DE BARI, Antonio; M., Rossella
Handle:
https://iris.unipv.it/handle/11571/713019
Published in:
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT